Acrel ARTM-8L Real-time Temperature Measuring Device
Specification
● Wireless temperature measurement, maximum measuring 8 points
● 5 digital outputs
● 3 analog output
● LCD
● Auxiliary power supply adapts with AC85~265V/DC100-300V
● 1 RS485 serial communication, Modbus-RTU
- Number of Channels Measured: 8
- Input Signal:
- NTC (Negative Temperature Coefficient Thermistors):
- R25 = 10.00kΩ; B25/50 = 3380K
- R25 = 50.00kΩ; B25/50 = 4150K
- R25 = 100.00kΩ; B25/50 = 3950K
- R25 = 30.00kΩ; B25/50 = 3950K
- PT100 / PT1000 (Three-wire system)
- Thermocouple support: K, T, J, E, N types; 0–5V
- NTC (Negative Temperature Coefficient Thermistors):
- Measurement Range:
- NTC: -40°C to 140°C
- PT100: -200°C to 850°C (depends on PT100 model)
- PT1000: -200°C to 850°C (depends on PT1000 model)
- Thermocouple: -200°C to 1372°C (depends on thermocouple model)
- Accuracy Class (Standard Signal):
- NTC: ±1°C
- PT100/PT1000: -200°C to 200°C ±1°C; +200°C to 850°C ±1%
- Thermocouple: 1%, 0–5V: ±0.5%
- Auxiliary Power:
- Voltage: AC85–265V / DC100–300V
- Power: ≤ 2W
- Alarm Output:
- 5 groups of normally open channels
- Capacity: AC250V/5A, DC30V/5A
- Temperature and Humidity Input:
- 1 optional channel
- Temperature: ±1°C
- Humidity: ±3%
- Transmit Output:
- 3 channels
- Transmission range: 4–20mA
- Communication:
- Protocol: MODBUS-RTU
- Interface: RS485
- Baud Rate: 2400, 4800, 9600, 19200
- Operating Environment:
- Temperature: -20°C to 55°C
- Humidity: ≤ 95%
Characteristic |
||
Measure 8 channels Pt100 temperature, it can also block any of these channels. | LED display | Easy wiring |
Dimension
Appearance
There are no reviews yet.